Vertical Laminar Flow Hood

Vertical Laminar Flow Hood

extractor hood
  1. Suctioned air
  2. Prefilter
  3. Pre-filtered air
  4. HEPA filter
  5. Motor fan
  6. 6 Sterile laminar flow in class 100 (or ISO 5)

The Economy vertical laminar flow hood mod. 700, was designed and created to allow manipulation of various materials in a sterile environment. It is particularly suitable for the PCR process. Total protection from germs and pulviscolo extends the use of this booth to the most varied sectors: microbiological, bacteriological, pharmaceutical, electronic, optical. The use of top quality materials and filters (absolute Hepa filter in class H14) in construction of this hood, allow it to meet the requirements of a class 100 (or ISO 5). The cabin must not be used when the treated material can represent a potential danger (pathogenic materials) for the operator and the environment.

The electrical components, the wiring of the switches and the electrical sockets located inside the control panel are isolated from the air flow to avoid inconveniences if handling flammable materials is carried out inside the cabin.

Information request

To enquire about this product please fill in the following form, our staff will get back to you as soon as possible.

Request a quote

To enquire about this product please fill in the following form, our staff will get back to you as soon as possible.

  • Construction in anti-acid powder coated steel
  • Extractable worktop in satin AISI 3042B stainless steel
  • Internal rear panel in satin AISI 304 2B stainless steel
  • < li>Side panels in 5 mm thick tempered glass.
  • Working area dimensions WxDxH: 700 x 410 x 490 mm.
  • Front opening 290 h.mm
  • Overall dimensions WxDxH: 770 x 660 x 800 mm.
  • Weight: 55 Kg.
  • Average air speed 0.40 / 0.45 m / sec.< /li>
  • Volume of filtered air 210 m3 / h.
  • Removable pre-filter in class G4 synthetic material (90% efficiency) acc. EN 779
  • Absolute Hepa filter (High efficiency particulate air) MPPS tested according to CEN 1822 regulations with global efficiency
  • 99.995% class H14, in class 100 at 0.3 micron, according to Fed Std 209E (Laser Test Royco 256) or ISO class
  • 5 according to ISO 14644.1.
  • ON/OFF switch for electric fan.
  • ON/OFF switch for lighting lamp.< /li>
  • ON/OFF switch for external service socket.
  • Low background noise electric fan, with the possibility of regulating the speed of the air flow, compliant with
  • < li>EN60335-1, EN 50178, EN 60950; VDE, CE, UL approvals.
  • Noise level ≤ 60dB
  • Electrical power supply: 230 V – 50 Hz.
  • Absorption: 140 W + 440 W
  • Protection fuses: 2 x 3 AF (5 X 20) mm fuses.
  • Power socket on the control panel protected with 2 AF (5 li>Mains connection socket: 10 A
  • Lighting with 18 W fluorescent lamp (1150 LUMEN) NB SAFETY

Vertical Laminar Flow Hood